Precise Measurement of Isotope Ratios with a Double-Focusing Magnetic Sector ICP Mass Spectrometer
Frank Vanhaecke, Luc Moëns, R. Dams, Philip Taylor
Abstract
Open-access reader
Frank Vanhaecke, Luc Moëns, R. Dams, Philip Taylor
Abstract
Open-access reader
The potential of a commercially available double-focusing magnetic sector ICP mass spectrometer (Element, Finnigan MAT, Bremen, Germany) for precise isotope ratio measurement at the low-resolution setting (R = 300) was evaluated. Optimization of scanning conditions led to a relative standard deviation for a set of 10 consecutive 2 min measurements of ∼0.1% ((206)Pb(+)/(207)Pb(+)) at signal intensities of ∼200 000 counts/s (peak height). This compares favorably with the best values ever reported for quadrupole ICPMS and barely exceeds the theoretical value (counting statistics). Increasing the signal intensity to values ≥500 000 counts/s (peak height) resulted in a further reduction of the RSDs obtained (for both (25)Mg(+)/(26)Mg(+) and (206)Pb(+)/(207)Pb(+)) to typically 0.04%. These figures are remarkably better than those reported for commercially available quadrupole ICPMS systems. This improvement significantly reduces the difference between isotope ratio precision of ICPMS on one hand and those of thermal ionization mass spectrometry and plasma source multiple collector mass spectrometry on the other.
OpenAlex reports 109 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
The potential of a commercially available double-focusing magnetic sector ICP mass spectrometer (Element, Finnigan MAT, Bremen, Germany) for precise isotope ratio measurement at the low-resolution setting (R = 300) was evaluated. Optimization of scanning conditions led to a relative standard deviation for a set of 10 consecutive 2 min measurements of ∼0.1% ((206)Pb(+)/(207)Pb(+)) at signal intensities of ∼200 000 counts/s (peak height). This compares favorably with the best values ever reported for quadrupole ICPMS and barely exceeds the theoretical value (counting statistics). Increasing the signal intensity to values ≥500 000 counts/s (peak height) resulted in a further reduction of the RSDs obtained (for both (25)Mg(+)/(26)Mg(+) and (206)Pb(+)/(207)Pb(+)) to typically 0.04%. These figures are remarkably better than those reported for commercially available quadrupole ICPMS systems. This improvement significantly reduces the difference between isotope ratio precision of ICPMS on one hand and those of thermal ionization mass spectrometry and plasma source multiple collector mass spectrometry on the other.
Key concepts: Chemistry, Isotope, Mass spectrometry, Analytical Chemistry (journal), Quadrupole mass analyzer, Thermal ionization mass spectrometry, Quadrupole, Inductively coupled plasma mass spectrometry