A Versatile Particle Beam Interface for Coupling Hplc to Ion Trap, Quadrupole and Sector Mass Spectrometers
Colin S. Creaser, James W. Stygall
Abstract
Colin S. Creaser, James W. Stygall
Abstract
A two-stage particle beam LC-MS interface, capable of being coupled to a quadrupole, sector or ion trap mass spectrometer, is described. The device has a modular design, allowing operating parameters to be easily optimized whilst in use. Successful coupling to a quadrupole ion trap mass spectrometer (ITMS) is achieved using a novel heated transfer line for analyte vaporization and scan routines which eliminate solvent ion effects.
OpenAlex reports 3 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
A two-stage particle beam LC-MS interface, capable of being coupled to a quadrupole, sector or ion trap mass spectrometer, is described. The device has a modular design, allowing operating parameters to be easily optimized whilst in use. Successful coupling to a quadrupole ion trap mass spectrometer (ITMS) is achieved using a novel heated transfer line for analyte vaporization and scan routines which eliminate solvent ion effects.
Key concepts: Ion trap, Quadrupole ion trap, Hybrid mass spectrometer, Mass spectrometry, Quadrupole mass analyzer, Quadrupole, Ion, Spectrometer