In vivo gamma measurements at very low levels with 4Π liquid scintillation detectors

Ernest C. Anderson

Open publisher page 13 citations

Abstract

A large 4Π liquid scintillation counter for the rapid measurement of total internal gamma contamination of the human body at concentrations 1000 times less than the maximum permissible levels is described. A counting time of only 100 seconds per subject is required to obtain a sensitivity of 5 × 10-10curies for 1.5 mev gamma rays. Because of the high counting efficiency and the great stability of operation, the magnitude of counter background is of comparatively little importance so that elaborate shielding precautions are not required. Energy resolution is sufficient to distinguish between gamma rays of markedly different energy, but not adequate for the identification of individual nuclides. The operating characteristics of the counter are discussed including the spectrum and stability of background, energy resolution, and the spectra and limits of sensitivity for nuclides of interest in connection with nuclear power stations and uranium mining operations.

About this research paper

What this paper is about

A large 4Π liquid scintillation counter for the rapid measurement of total internal gamma contamination of the human body at concentrations 1000 times less than the maximum permissible levels is described. A counting time of only 100 seconds per subject is required to obtain a sensitivity of 5 × 10-10curies for 1.5 mev gamma rays. Because of the high counting efficiency and the great stability of operation, the magnitude of counter background is of comparatively little importance so that elaborate shielding precautions are not required. Energy resolution is sufficient to distinguish between gamma rays of markedly different energy, but not adequate for the identification of individual nuclides. The operating characteristics of the counter are discussed including the spectrum and stability of background, energy resolution, and the spectra and limits of sensitivity for nuclides of interest in connection with nuclear power stations and uranium mining operations.

Why it matters

OpenAlex reports 13 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

A large 4Π liquid scintillation counter for the rapid measurement of total internal gamma contamination of the human body at concentrations 1000 times less than the maximum permissible levels is described. A counting time of only 100 seconds per subject is required to obtain a sensitivity of 5 × 10-10curies for 1.5 mev gamma rays. Because of the high counting efficiency and the great stability of operation, the magnitude of counter background is of comparatively little importance so that elaborate shielding precautions are not required. Energy resolution is sufficient to distinguish between gamma rays of markedly different energy, but not adequate for the identification of individual nuclides. The operating characteristics of the counter are discussed including the spectrum and stability of background, energy resolution, and the spectra and limits of sensitivity for nuclides of interest in connection with nuclear power stations and uranium mining operations.

Key concepts: Nuclide, Scintillation counter, Liquid scintillation counting, Scintillation, Physics, Gamma spectroscopy, Nuclear physics, Gamma ray

Related papers

Back to paper searchBrowse research topicsOriginal source
In vivo gamma measurements at very low levels with 4Π liquid scintillation detectors — Research Paper | ScholarLens