Response of Rotary Blood Pumps to Changes in Preload and Afterload at a Fixed Speed Setting Are Unphysiological When Compared With the Natural Heart

Robert F. Salamonsen, D. G. Mason, Peter Ayre

Open publisher page 147 citations

Abstract

Responses of four rotary blood pumps (Incor, Heartmate II, Heartware, and Duraheart) at a single speed setting to changes in preload and afterload were assessed using the human left ventricle as a benchmark for comparison. Data for the rotary pumps were derived from pressure flow relations reported in the literature while the natural heart was characterized by the Frank-Starling curve adjusted to fit outputs at different afterloads reported in the literature. Preload sensitivity (mean ± SD) for all pumps at all afterloads tested was 0.105 ± 0.092 L/min/mm Hg, while afterload sensitivity was 0.09 ± 0.034 L/min/mm Hg-values that were not significantly different (t-test, P = 0.56). By contrast, preload sensitivity of the natural heart was over twice as high (0.213 ± 0.03 L/min/mm Hg) and afterload sensitivity about one-third (0.03 ± 0.01 L/min/mm Hg) the values recorded for rotary pumps (t-test, P < 0.001). Maximum preload sensitivity and minimum afterload sensitivity allow the right and left ventricles to synchronize outputs without neural or humoral intervention. This theoretical study reinforces the need to provide preload sensitive control mechanisms of sufficient power to enable the pump and left ventricle in combination to adapt to changes in right ventricular output automatically.

About this research paper

What this paper is about

Responses of four rotary blood pumps (Incor, Heartmate II, Heartware, and Duraheart) at a single speed setting to changes in preload and afterload were assessed using the human left ventricle as a benchmark for comparison. Data for the rotary pumps were derived from pressure flow relations reported in the literature while the natural heart was characterized by the Frank-Starling curve adjusted to fit outputs at different afterloads reported in the literature. Preload sensitivity (mean ± SD) for all pumps at all afterloads tested was 0.105 ± 0.092 L/min/mm Hg, while afterload sensitivity was 0.09 ± 0.034 L/min/mm Hg-values that were not significantly different (t-test, P = 0.56). By contrast, preload sensitivity of the natural heart was over twice as high (0.213 ± 0.03 L/min/mm Hg) and afterload sensitivity about one-third (0.03 ± 0.01 L/min/mm Hg) the values recorded for rotary pumps (t-test, P < 0.001). Maximum preload sensitivity and minimum afterload sensitivity allow the right and left ventricles to synchronize outputs without neural or humoral intervention. This theoretical study reinforces the need to provide preload sensitive control mechanisms of sufficient power to enable the pump and left ventricle in combination to adapt to changes in right ventricular output automatically.

Why it matters

OpenAlex reports 147 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Responses of four rotary blood pumps (Incor, Heartmate II, Heartware, and Duraheart) at a single speed setting to changes in preload and afterload were assessed using the human left ventricle as a benchmark for comparison. Data for the rotary pumps were derived from pressure flow relations reported in the literature while the natural heart was characterized by the Frank-Starling curve adjusted to fit outputs at different afterloads reported in the literature. Preload sensitivity (mean ± SD) for all pumps at all afterloads tested was 0.105 ± 0.092 L/min/mm Hg, while afterload sensitivity was 0.09 ± 0.034 L/min/mm Hg-values that were not significantly different (t-test, P = 0.56). By contrast, preload sensitivity of the natural heart was over twice as high (0.213 ± 0.03 L/min/mm Hg) and afterload sensitivity about one-third (0.03 ± 0.01 L/min/mm Hg) the values recorded for rotary pumps (t-test, P < 0.001). Maximum preload sensitivity and minimum afterload sensitivity allow the right and left ventricles to synchronize outputs without neural or humoral intervention. This theoretical study reinforces the need to provide preload sensitive control mechanisms of sufficient power to enable the pump and left ventricle in combination to adapt to changes in right ventricular output automatically.

Key concepts: Preload, Afterload, Ventricle, Cardiology, Sensitivity (control systems), Hemodynamics, Internal medicine, Blood pressure