Response of Rotary Blood Pumps to Changes in Preload and Afterload at a Fixed Speed Setting Are Unphysiological When Compared With the Natural Heart
Robert F. Salamonsen, D. G. Mason, Peter Ayre
Abstract
Robert F. Salamonsen, D. G. Mason, Peter Ayre
Abstract
Responses of four rotary blood pumps (Incor, Heartmate II, Heartware, and Duraheart) at a single speed setting to changes in preload and afterload were assessed using the human left ventricle as a benchmark for comparison. Data for the rotary pumps were derived from pressure flow relations reported in the literature while the natural heart was characterized by the Frank-Starling curve adjusted to fit outputs at different afterloads reported in the literature. Preload sensitivity (mean ± SD) for all pumps at all afterloads tested was 0.105 ± 0.092 L/min/mm Hg, while afterload sensitivity was 0.09 ± 0.034 L/min/mm Hg-values that were not significantly different (t-test, P = 0.56). By contrast, preload sensitivity of the natural heart was over twice as high (0.213 ± 0.03 L/min/mm Hg) and afterload sensitivity about one-third (0.03 ± 0.01 L/min/mm Hg) the values recorded for rotary pumps (t-test, P < 0.001). Maximum preload sensitivity and minimum afterload sensitivity allow the right and left ventricles to synchronize outputs without neural or humoral intervention. This theoretical study reinforces the need to provide preload sensitive control mechanisms of sufficient power to enable the pump and left ventricle in combination to adapt to changes in right ventricular output automatically.
OpenAlex reports 147 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Responses of four rotary blood pumps (Incor, Heartmate II, Heartware, and Duraheart) at a single speed setting to changes in preload and afterload were assessed using the human left ventricle as a benchmark for comparison. Data for the rotary pumps were derived from pressure flow relations reported in the literature while the natural heart was characterized by the Frank-Starling curve adjusted to fit outputs at different afterloads reported in the literature. Preload sensitivity (mean ± SD) for all pumps at all afterloads tested was 0.105 ± 0.092 L/min/mm Hg, while afterload sensitivity was 0.09 ± 0.034 L/min/mm Hg-values that were not significantly different (t-test, P = 0.56). By contrast, preload sensitivity of the natural heart was over twice as high (0.213 ± 0.03 L/min/mm Hg) and afterload sensitivity about one-third (0.03 ± 0.01 L/min/mm Hg) the values recorded for rotary pumps (t-test, P < 0.001). Maximum preload sensitivity and minimum afterload sensitivity allow the right and left ventricles to synchronize outputs without neural or humoral intervention. This theoretical study reinforces the need to provide preload sensitive control mechanisms of sufficient power to enable the pump and left ventricle in combination to adapt to changes in right ventricular output automatically.
Key concepts: Preload, Afterload, Ventricle, Cardiology, Sensitivity (control systems), Hemodynamics, Internal medicine, Blood pressure