Desktop analysers: quality of results obtained by medical office personnel.

Amin A. Nanji, Raymond Poon, I Hinberg

Open full text 5 citations

Abstract

We carried out a study to evaluate the quality of results obtained by 14 nontechnical medical office personnel using desktop analysers. The instruments evaluated were the Reflotron analyser, the Seralyzer, the Vision analyser and the DT60 analyser. For precision studies low and high concentrations of control materials were used. For correlation studies the results obtained by the office personnel were compared with those obtained by a trained technologist. The coefficient of variation for the office personnel ranged from 3.0% to 8.1% with the Reflotron analyser, from 6.3% to 26.5% with the Seralyzer, from 1.0% to 4.1% with the Vision analyser and from 1.4% to 16.7% with the DT60 analyser. The correlation coefficient ranged from 0.970 to 0.997 with the Reflotron analyser, from 0.779 to 0.997 with the Seralyzer, from 0.975 to 0.998 with the Vision analyser and from 0.963 to 0.995 with the DT60 analyser. The proportion of results obtained by the office personnel that differed by more than 10% from those obtained by the technologist was 7% with the Reflotron analyser, 42% with the Seralyzer, 2% with the Vision analyser and 21% with the DT60 analyser. The instruments whose operation involves the least number of steps gave the most reliable results in the hands of medical office personnel.

About this research paper

What this paper is about

We carried out a study to evaluate the quality of results obtained by 14 nontechnical medical office personnel using desktop analysers. The instruments evaluated were the Reflotron analyser, the Seralyzer, the Vision analyser and the DT60 analyser. For precision studies low and high concentrations of control materials were used. For correlation studies the results obtained by the office personnel were compared with those obtained by a trained technologist. The coefficient of variation for the office personnel ranged from 3.0% to 8.1% with the Reflotron analyser, from 6.3% to 26.5% with the Seralyzer, from 1.0% to 4.1% with the Vision analyser and from 1.4% to 16.7% with the DT60 analyser. The correlation coefficient ranged from 0.970 to 0.997 with the Reflotron analyser, from 0.779 to 0.997 with the Seralyzer, from 0.975 to 0.998 with the Vision analyser and from 0.963 to 0.995 with the DT60 analyser. The proportion of results obtained by the office personnel that differed by more than 10% from those obtained by the technologist was 7% with the Reflotron analyser, 42% with the Seralyzer, 2% with the Vision analyser and 21% with the DT60 analyser. The instruments whose operation involves the least number of steps gave the most reliable results in the hands of medical office personnel.

Why it matters

OpenAlex reports 5 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

We carried out a study to evaluate the quality of results obtained by 14 nontechnical medical office personnel using desktop analysers. The instruments evaluated were the Reflotron analyser, the Seralyzer, the Vision analyser and the DT60 analyser. For precision studies low and high concentrations of control materials were used. For correlation studies the results obtained by the office personnel were compared with those obtained by a trained technologist. The coefficient of variation for the office personnel ranged from 3.0% to 8.1% with the Reflotron analyser, from 6.3% to 26.5% with the Seralyzer, from 1.0% to 4.1% with the Vision analyser and from 1.4% to 16.7% with the DT60 analyser. The correlation coefficient ranged from 0.970 to 0.997 with the Reflotron analyser, from 0.779 to 0.997 with the Seralyzer, from 0.975 to 0.998 with the Vision analyser and from 0.963 to 0.995 with the DT60 analyser. The proportion of results obtained by the office personnel that differed by more than 10% from those obtained by the technologist was 7% with the Reflotron analyser, 42% with the Seralyzer, 2% with the Vision analyser and 21% with the DT60 analyser. The instruments whose operation involves the least number of steps gave the most reliable results in the hands of medical office personnel.

Key concepts: Analyser, Medicine, Computer science, Chromatography, Chemistry

Related papers

Back to paper searchBrowse research topicsOriginal source
Desktop analysers: quality of results obtained by medical office personnel. — Research Paper | ScholarLens