1996Clinical ChemistryRequires access

Quantification of vanadium in serum by electrothermal atomic absorption spectrometry

G. Heinemann, Wolfgang Vogt

Open publisher page 29 citations

Abstract

We describe an ultrasensitive and reliable method for determining vanadium in human serum by electrothermal atomic absorption spectrometry. After lyophilization, the serum is digested in acid at high pressure, and the digests are evaporated to a small volume. Vanadium in the digests is complexed with cupferron, extracted, and dried. The residue is redissolved in formic acid, where it is 15-fold more concentrated than in the original serum sample. To enhance the furnace sensitivity, we injected six 40-microL aliquots (total, 240 microL) of the concentrated extract. The median concentration of vanadium in 108 persons was 50 ng/L, in good agreement with previously reported results by neutron activation analysis. The characteristic mass obtained (the mass required to give a signal of 0.0044A. s) was 28 pg, the limit of detection 11 ng/L, the limit of quantification 17 ng/L, and the total imprecision (CV) 5.5% at 1.54 micrograms/L. In two assays of Standard Reference Material 1577a (certified vanadium content 99 +/- 8 ng/g), we obtained values of 94.1 and 97 ng/g.

About this research paper

What this paper is about

We describe an ultrasensitive and reliable method for determining vanadium in human serum by electrothermal atomic absorption spectrometry. After lyophilization, the serum is digested in acid at high pressure, and the digests are evaporated to a small volume. Vanadium in the digests is complexed with cupferron, extracted, and dried. The residue is redissolved in formic acid, where it is 15-fold more concentrated than in the original serum sample. To enhance the furnace sensitivity, we injected six 40-microL aliquots (total, 240 microL) of the concentrated extract. The median concentration of vanadium in 108 persons was 50 ng/L, in good agreement with previously reported results by neutron activation analysis. The characteristic mass obtained (the mass required to give a signal of 0.0044A. s) was 28 pg, the limit of detection 11 ng/L, the limit of quantification 17 ng/L, and the total imprecision (CV) 5.5% at 1.54 micrograms/L. In two assays of Standard Reference Material 1577a (certified vanadium content 99 +/- 8 ng/g), we obtained values of 94.1 and 97 ng/g.

Why it matters

OpenAlex reports 29 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

We describe an ultrasensitive and reliable method for determining vanadium in human serum by electrothermal atomic absorption spectrometry. After lyophilization, the serum is digested in acid at high pressure, and the digests are evaporated to a small volume. Vanadium in the digests is complexed with cupferron, extracted, and dried. The residue is redissolved in formic acid, where it is 15-fold more concentrated than in the original serum sample. To enhance the furnace sensitivity, we injected six 40-microL aliquots (total, 240 microL) of the concentrated extract. The median concentration of vanadium in 108 persons was 50 ng/L, in good agreement with previously reported results by neutron activation analysis. The characteristic mass obtained (the mass required to give a signal of 0.0044A. s) was 28 pg, the limit of detection 11 ng/L, the limit of quantification 17 ng/L, and the total imprecision (CV) 5.5% at 1.54 micrograms/L. In two assays of Standard Reference Material 1577a (certified vanadium content 99 +/- 8 ng/g), we obtained values of 94.1 and 97 ng/g.

Key concepts: Vanadium, Chemistry, Detection limit, Atomic absorption spectroscopy, Mass spectrometry, Certified reference materials, Chromatography, Formic acid

Related papers

Back to paper searchBrowse research topicsOriginal source
Quantification of vanadium in serum by electrothermal atomic absorption spectrometry — Research Paper | ScholarLens