Monte Carlo radiography simulation for assessment of absorbed radiation dose in femur bone marrow during X-ray radiography for constant mAs and AEC techniques
Zaker Salehi, Wan Ahmad Kamil, Biswa Mohan Biswal, Ahmad Lutfi Yusoff
Abstract
Zaker Salehi, Wan Ahmad Kamil, Biswa Mohan Biswal, Ahmad Lutfi Yusoff
Abstract
Background : The purpose of this study was to find the accurate absorbed dose in the femur bone marrow during the X-ray radiography for constant mAs and AEC techniques. Materials and Methods : The DOSXYZnrc was used to simulate radia$on doses in two human femurs during diagnos$c radiography. EGSnrc phantoms produced from actual CT images of human femurs were modified by adding seven micrometre layers of marrow $ssues . The X-ray machine was simulated using BEAMnrc using 30 billions par$cles for different combina$ons of energies and filters. The resultant data was used to in DOSXYSnrc simula$ons to evaluate the absorbed dose in the human femur. Results : In the head of the femur, for 2.5 mm aluminium filtered 85 kVp X-ray set at 50 mAs, the absorbed dose in the marrow was found to be 1.360 mGy, ~ 36% of the absorbed dose in the cor$cal bone. It was also found that for the constant mAs technique, the radia$on dose in the marrow over the studied energies and filter combina$on, ranges from 0.356 mGy to 2.403 mGy, with higher dose recorded for higher kVp se8ngs. However, for the AE C technique, the dose is lower for higher kVp se8ngs. For a typical se8ng, v iz. 85 kVp, 6 mAs at 48 inches SID, the bone marrow absorbed dose was found to be 0.186 mGy for the constant mAs technique and 0.0308 mGy for the AEC technique. Conclusion : It was confirmed that the radia$on dose is lower when the AEC exposure technique is used as opposed to using constant mAs technique. For the AEC technique, typical dose to the bone marrow was found to be ~ 0.05 mGy, decreasing with both kVp se8ngs and beam filtra $on. For constant mAs technique, the typical dose to bone marrow is found to be higher, ~ 0.2 mGy, decreasing with the amount of filtra$on used but increasing with the kVp se8ng.
OpenAlex reports 2 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Background : The purpose of this study was to find the accurate absorbed dose in the femur bone marrow during the X-ray radiography for constant mAs and AEC techniques. Materials and Methods : The DOSXYZnrc was used to simulate radia$on doses in two human femurs during diagnos$c radiography. EGSnrc phantoms produced from actual CT images of human femurs were modified by adding seven micrometre layers of marrow $ssues . The X-ray machine was simulated using BEAMnrc using 30 billions par$cles for different combina$ons of energies and filters. The resultant data was used to in DOSXYSnrc simula$ons to evaluate the absorbed dose in the human femur. Results : In the head of the femur, for 2.5 mm aluminium filtered 85 kVp X-ray set at 50 mAs, the absorbed dose in the marrow was found to be 1.360 mGy, ~ 36% of the absorbed dose in the cor$cal bone. It was also found that for the constant mAs technique, the radia$on dose in the marrow over the studied energies and filter combina$on, ranges from 0.356 mGy to 2.403 mGy, with higher dose recorded for higher kVp se8ngs. However, for the AE C technique, the dose is lower for higher kVp se8ngs. For a typical se8ng, v iz. 85 kVp, 6 mAs at 48 inches SID, the bone marrow absorbed dose was found to be 0.186 mGy for the constant mAs technique and 0.0308 mGy for the AEC technique. Conclusion : It was confirmed that the radia$on dose is lower when the AEC exposure technique is used as opposed to using constant mAs technique. For the AEC technique, typical dose to the bone marrow was found to be ~ 0.05 mGy, decreasing with both kVp se8ngs and beam filtra $on. For constant mAs technique, the typical dose to bone marrow is found to be higher, ~ 0.2 mGy, decreasing with the amount of filtra$on used but increasing with the kVp se8ng.
Key concepts: Nuclear medicine, Radiography, Absorbed dose, X-ray, Materials science, Femur, Radiation dose, Bone marrow