Tissue Harmonic Imaging
Kazuaki Tanabe, Marek Bëlohlávek, James F. Greenleaf, James B. Seward
Abstract
Open-access reader
Kazuaki Tanabe, Marek Bëlohlávek, James F. Greenleaf, James B. Seward
Abstract
Open-access reader
Tissue harmonic scanning visually improves echocardiographic image quality. The aim of the present study was to objectively assess the improvement in harmonic image quality under controlled laboratory conditions. A tissue-mimicking phantom that contained 8-mm-diameter cystic lesions at depths ranging from 2 to 12 cm was used. Harmonic scans (1.7 MHz transmit, 3.4 MHz receive) of the phantom were obtained and lesion detectability was compared to that in scans acquired with 2 fundamental frequencies (2.0 and 3.3 MHz). A 2 cm-thick ethanol layer was also used to simulate the nonlinear effect of human fat. Cyst detectability was quantified by measurement of the contrast-to-speckle ratio (CSR). The results indicated no significant difference in the CSR between harmonic and fundamental images obtained without the ethanol layer. With images obtained with the ethanol layer, a relative increase of the CSR during harmonic imaging was observed with respect to fundamental imaging (p<0.05). In conclusion, a fat layer, here simulated by ethanol, plays a significant role in determining the resulting image quality. Without this layer, the contribution of the second harmonic mode was not significant. Thus, in a slim patient, the harmonic mode may not be as beneficial to image improvement as in an obese patient.
OpenAlex reports 21 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Tissue harmonic scanning visually improves echocardiographic image quality. The aim of the present study was to objectively assess the improvement in harmonic image quality under controlled laboratory conditions. A tissue-mimicking phantom that contained 8-mm-diameter cystic lesions at depths ranging from 2 to 12 cm was used. Harmonic scans (1.7 MHz transmit, 3.4 MHz receive) of the phantom were obtained and lesion detectability was compared to that in scans acquired with 2 fundamental frequencies (2.0 and 3.3 MHz). A 2 cm-thick ethanol layer was also used to simulate the nonlinear effect of human fat. Cyst detectability was quantified by measurement of the contrast-to-speckle ratio (CSR). The results indicated no significant difference in the CSR between harmonic and fundamental images obtained without the ethanol layer. With images obtained with the ethanol layer, a relative increase of the CSR during harmonic imaging was observed with respect to fundamental imaging (p<0.05). In conclusion, a fat layer, here simulated by ethanol, plays a significant role in determining the resulting image quality. Without this layer, the contribution of the second harmonic mode was not significant. Thus, in a slim patient, the harmonic mode may not be as beneficial to image improvement as in an obese patient.
Key concepts: Second-harmonic imaging microscopy, Imaging phantom, Harmonic, Image quality, Materials science, Biomedical engineering, Speckle pattern, Optics