Non-cycloplegic spherical equivalent refraction in adults: comparison of the double-pass system, retinoscopy, subjective refraction and a table-mounted autorefractor.

Meritxell Vilaseca, Montserrat Arjona, Jaume Pujol, Elvira Peris, Vanessa Martínez

Open full text 9 citations

Abstract

AIM: To evaluate the accuracy of spherical equivalent (SE) estimates of a double-pass system and to compare it with retinoscopy, subjective refraction and a table-mounted autorefractor. METHODS: Non-cycloplegic refraction was performed on 125 eyes of 65 healthy adults (age 23.5±3.0 years) from October 2010 to January 2011 using retinoscopy, subjective refraction, autorefraction (Auto kerato-refractometer TOPCON KR-8100, Japan) and a double-pass system (Optical Quality Analysis System, OQAS, Visiometrics S.L., Spain). Nine consecutive measurements with the double-pass system were performed on a subgroup of 22 eyes to assess repeatability. To evaluate the trueness of the OQAS instrument, the SE laboratory bias between the double-pass system and the other techniques was calculated. RESULTS: The SE mean coefficient of repeatability obtained was 0.22D. Significant correlations could be established between the OQAS and the SE obtained with retinoscopy (r=0.956, P<0.001), subjective refraction (r=0.955, P<0.001) and autorefraction (r=0.957, P<0.001). The differences in SE between the double-pass system and the other techniques were significant (P<0.001), but lacked clinical relevance except for retinoscopy; Retinoscopy gave more hyperopic values than the double-pass system -0.51±0.50D as well as the subjective refraction -0.23±0.50D; More myopic values were achieved by means of autorefraction 0.24±0.49D. CONCLUSION: The double-pass system provides accurate and reliable estimates of the SE that can be used for clinical studies. This technique can determine the correct focus position to assess the ocular optical quality. However, it has a relatively small measuring range in comparison with autorefractors (-8.00 to +5.00D), and requires prior information on the refractive state of the patient.

About this research paper

What this paper is about

AIM: To evaluate the accuracy of spherical equivalent (SE) estimates of a double-pass system and to compare it with retinoscopy, subjective refraction and a table-mounted autorefractor. METHODS: Non-cycloplegic refraction was performed on 125 eyes of 65 healthy adults (age 23.5±3.0 years) from October 2010 to January 2011 using retinoscopy, subjective refraction, autorefraction (Auto kerato-refractometer TOPCON KR-8100, Japan) and a double-pass system (Optical Quality Analysis System, OQAS, Visiometrics S.L., Spain). Nine consecutive measurements with the double-pass system were performed on a subgroup of 22 eyes to assess repeatability. To evaluate the trueness of the OQAS instrument, the SE laboratory bias between the double-pass system and the other techniques was calculated. RESULTS: The SE mean coefficient of repeatability obtained was 0.22D. Significant correlations could be established between the OQAS and the SE obtained with retinoscopy (r=0.956, P<0.001), subjective refraction (r=0.955, P<0.001) and autorefraction (r=0.957, P<0.001). The differences in SE between the double-pass system and the other techniques were significant (P<0.001), but lacked clinical relevance except for retinoscopy; Retinoscopy gave more hyperopic values than the double-pass system -0.51±0.50D as well as the subjective refraction -0.23±0.50D; More myopic values were achieved by means of autorefraction 0.24±0.49D. CONCLUSION: The double-pass system provides accurate and reliable estimates of the SE that can be used for clinical studies. This technique can determine the correct focus position to assess the ocular optical quality. However, it has a relatively small measuring range in comparison with autorefractors (-8.00 to +5.00D), and requires prior information on the refractive state of the patient.

Why it matters

OpenAlex reports 9 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

AIM: To evaluate the accuracy of spherical equivalent (SE) estimates of a double-pass system and to compare it with retinoscopy, subjective refraction and a table-mounted autorefractor. METHODS: Non-cycloplegic refraction was performed on 125 eyes of 65 healthy adults (age 23.5±3.0 years) from October 2010 to January 2011 using retinoscopy, subjective refraction, autorefraction (Auto kerato-refractometer TOPCON KR-8100, Japan) and a double-pass system (Optical Quality Analysis System, OQAS, Visiometrics S.L., Spain). Nine consecutive measurements with the double-pass system were performed on a subgroup of 22 eyes to assess repeatability. To evaluate the trueness of the OQAS instrument, the SE laboratory bias between the double-pass system and the other techniques was calculated. RESULTS: The SE mean coefficient of repeatability obtained was 0.22D. Significant correlations could be established between the OQAS and the SE obtained with retinoscopy (r=0.956, P<0.001), subjective refraction (r=0.955, P<0.001) and autorefraction (r=0.957, P<0.001). The differences in SE between the double-pass system and the other techniques were significant (P<0.001), but lacked clinical relevance except for retinoscopy; Retinoscopy gave more hyperopic values than the double-pass system -0.51±0.50D as well as the subjective refraction -0.23±0.50D; More myopic values were achieved by means of autorefraction 0.24±0.49D. CONCLUSION: The double-pass system provides accurate and reliable estimates of the SE that can be used for clinical studies. This technique can determine the correct focus position to assess the ocular optical quality. However, it has a relatively small measuring range in comparison with autorefractors (-8.00 to +5.00D), and requires prior information on the refractive state of the patient.

Key concepts: Autorefractor, Retinoscopy, Subjective refraction, Repeatability, Mathematics, Refraction, Optometry, Theodolite

Related papers

Back to paper searchBrowse research topicsOriginal source
Non-cycloplegic spherical equivalent refraction in adults: comparison of the double-pass system, retinoscopy, subjective refraction and a table-mounted autorefractor. — Research Paper | ScholarLens